⬇ Cite as BibTeX (.bib) · ⬇ CSL-JSON (Zotero / Pandoc) · ⬇ RIS (EndNote) · blake3:f52f3e2ddf373400

Encoding unit — one grounded idea

1 EWOD electrode array reference (80-electrode, 1.5 mm pitch): 80 individually addressable square electrodes at 1.5 mm × 1.5 mm pitch arranged in 8 × 10 array; reservoir electrodes (3 mm × 3 mm) at four corners; gap between electrodes 50 µm; total active area 12 mm × 15 mm. Top plate is unpatterned ITO common ground.
“80 individually addressable square electrodes at 1.5 mm × 1.5 mm pitch arranged in 8 × 10 array; reservoir electrodes (3 mm × 3 mm) at four corners; gap between electrodes 50 µm; total active area 12 mm × 15 mm. Top plate is unpatterned ITO”
written · p. 1 · 2026-05-09 · en · OpenIE Microfluidics Commons · EWOD electrode array reference (80-electrode, 1.5 mm pitch)
id: 9027c17e2bd8189a

from the paper Free Microfluidics Open Engineering Suite: Validated Fabrication Recipes, Failure Modes, CAD Designs & Instruments (CC0-1.0) (2026)

concepts in this idea